Specifications
| Technical data | |
|---|---|
| Measuring quantity | Standard flow velocity wN based on standard conditions 1) Temperature of medium TM |
| Measured medium | Air or nitrogen, other gases upon request |
| Measuring range wN | 0 … 1 / 10 / 20 m/s; Special measuring range inbetween 1 … 20 m/s (steps of 0.1 m/s) |
| Lower detection limit wN | 0.06 m/s |
| Measuring range TM | -20 … +70 °C |
| Measuring accuracy | |
| Standard wN | ±(5 % of measured value + [0.4 % of fmr; min 0.02 m/s]) 2) |
| High precision (optional) wN | ±(3 % of measured value + [0.4 % of fmr; min 0.02 m/s]) 2) |
| Response time (t90) wN | 3 s (jump from 0 to 5 m/s in air) |
| Temperature gradient wN | < 2 K/min (at wN = 5 m/s in air) |
| TM (for wN > 1 m/s) | ±1 K (10 … 30 °C); ±2 K (remaining measuring range) |
| Operating temperature | |
| Sensor / electronics | -20 … +70 °C |
| Storage temperature | -30 … +85 °C |
| General data | |
| Humidity | Measuring mode: Non-condensing (< 95 % RH) |
| Operating pressure | Atmospheric (700 … 1,300 hPa) |
| Display | Dual-LED: - Green: Status - Red (flashing): Sensor defective |
| Supply voltage | 24 V AC/DC ± 10 % |
| Current consumption | Typical < 60 mA, max. 100 mA |
| Analog outputs Type: Auto U / I) | wN, TM 0 … 10 V / 4 … 20 mA (switching of signal characteristic depends on RL) -Voltage output (0 … 10 V): RL ≥ 550 Ω (short-circuit protected) - Current output (4 … 20 mA): RL ≤ 500 Ω - Change-over hysteresis: 50 Ω - CL ≤ 10 nF Voltage output: RL ≥ 500 Ω Current output: RL ≤ 500 Ω Hysteresis: 50 Ω |
| Electrical connection | Non-detachable cable, 5 x 0,25 mm2 Standard: Length 2 m or selectable (3 … 100 m; steps of 1 m), with ferrules Hygienic Design Option: 1.5 m, M12 plug (male, A-coded, 5 pole); Additional content of delivery: Cable 10 m, M12 bushing (female, A-coded, 5 pole) |
| Maximum cable length | Voltage interface / current interface: 100 m Voltage mode: Use AGND and RL ≥ 10 k (mass offset) |
| Mounting position | Arbitrary |
| Minimum immersion depth | 58 mm (< 58 mm upon request) |
| Protection type | IP65 |
| Protection class | III (SELV or PELV) |
| Probe length | 300/500 mm |
| Weight | 200 g max. |
| Material | |
| Sensor tube | Stainless steel 1.4571 |
| Sensor head | PBT (glass-fibre reinforced), stainless steel 1.4571 |
| Protective coating (optional) | Polyurethan / parylene (coating not available for Hygienic Design version) |
| Sensor cable | Standard: PUR, halogen-free Hygienic Design Option: PP-EPDM, halogen-free |
1) Standard conditions: TN = 20 °C and pN = 1,013.25 hPa
2) Related to adjustment reference; "fmr" = final measuring range
Subject to technical changes.